Metodología de evaluación de la incidencia del factor humano en la medición intra-laboratorio de volúmenes de líquidos mediante el método gravimétrico y de transferencia en recipientes volumétricos metálicos.
dc.contributor.advisor | Bermudez Rubio, Dagoberto | |
dc.contributor.author | Aguirre Romero, Elvis | |
dc.coverage.spatial | Bogotá - Colombia | spa |
dc.creator.email | dccaceresr@libertadores.edu.co | spa |
dc.date.accessioned | 2023-04-28T02:46:53Z | |
dc.date.available | 2023-04-28T02:46:53Z | |
dc.date.created | 2023 | |
dc.description | Los Laboratorios de la Superintendencia de Industria y Comercio (SIC) implementan estudios de verificación intra-laboratorio para demostrar la confiabilidad de los métodos de calibración de recipientes volumétricos Metálicos (RVM) utilizados en las actividades de inspección y de verificación de control metrológico. La diversidad de métodos de medición de volumen de líquidos y de instrumentos volumétricos empleados genera la necesidad de soportar con técnicas estadísticas que los resultados obtenidos por el método de medición gravimétrico y de trasferencia son equiparables. El objetivo de esta investigación es evaluar la incidencia del factor humano en la medición intra-Laboratorio de volumen de líquidos mediante los métodos gravimétrico y de transferencia en RVM con capacidad de 20 Litros. Se consideró un diseño experimental de dos factores de naturaleza nominal con 3 y 2 niveles respectivamente. A partir del diseño se construyó el arreglo factorial completo 3x2 con cinco replicas, que permitirá obtener toda la información relevante en relación al efecto de estos factores sobre las mediciones de volumen total. La prueba ANOVA de dos factores muestra que no hay diferencias significativas entre los promedios de las mediciones volumétricas por los dos tratamientos: analistas y métodos de medición. El modelo estadístico propuesto permite dar trazabilidaddelainfluenciadelfactorhumanoenlamedicióndevolúmenesdelíquidosenRVMpormétodosdiferentes. | spa |
dc.description.abstract | TheSuperintendenciaofIndustriaandComercio(SIC)CalibrationLaboratoriesimplementintra-laboratoryverification studies to demonstrate the reliability of the calibration methods for metallic volumetric containers (RVM) used in the inspection and verification activities of metrological control. The diversity of liquid volume measurement methods and volumetric instruments used generates the need to support with statistical techniques that the results obtained by the gravimetric and transfer measurement method are comparable. The objective of this research is to evaluate the incidence of the human factor in the intra-laboratory measurement of liquid volume by gravimetric and transfer methods in RVM with a capacity of 20 liters. An experimental design of two nominal factors with 3 and 2 levels respectively was considered. From the design,was constructed the complete 3x2 factorial arrangement with five replicates , which will allow obtaining all the relevant information regarding the effect of these factors on the total volume measurements. The two-factor ANOVA test shows that there are no significant differences between the means of the volumetric measurements by the two treatments: analysts and measurement methods . The proposed statistical model allows traceability of the influence of the human factor in the measurement of volumes of liquids in RVM by different methods. | spa |
dc.format | spa | |
dc.identifier.uri | http://hdl.handle.net/11371/5755 | |
dc.language.iso | spa | spa |
dc.publisher | Fundación Universitaria Los Libertadores. Sede Bogotá. | spa |
dc.rights.acceso | Abierto (Texto Completo) | spa |
dc.rights.accessrights | OpenAccess | spa |
dc.subject.lemb | Estadística - Métodos gráficos | spa |
dc.subject.lemb | Laboratorios de hidráulica | spa |
dc.subject.lemb | Análisis estadistico | spa |
dc.subject.lemb | Instrumentos de medición | spa |
dc.subject.proposal | ANOVA de dos factores | spa |
dc.subject.proposal | Medición volumétrica de líquidos | spa |
dc.subject.proposal | Factor humano | spa |
dc.subject.proposal | Menisco. | spa |
dc.subject.subjectenglish | Two-factor ANOVA | spa |
dc.subject.subjectenglish | Volumetric measurement of liquids | spa |
dc.subject.subjectenglish | Human factor | spa |
dc.subject.subjectenglish | Meniscus | spa |
dc.title | Metodología de evaluación de la incidencia del factor humano en la medición intra-laboratorio de volúmenes de líquidos mediante el método gravimétrico y de transferencia en recipientes volumétricos metálicos. | spa |
dc.title.titleenglish | Methodology for the evaluation of the incidence of the human factor in the intra-laboratory measurement of liquid volumes by the gravimetric and transfer method in metallic volumetric vessels. | spa |
dc.type | info:eu-repo/semantics/bachelorThesis | spa |
Archivos
Bloque original
1 - 1 de 1
Cargando...
- Nombre:
- Aguirre_Elvis_2023.pdf
- Tamaño:
- 1.17 MB
- Formato:
- Adobe Portable Document Format
- Descripción: